Innovative Designs Inc 10KSB, filed Jan 31, 2007: Test Results from Texas Research Institute Austin, Inc.
EXHIBIT 99.2
PERMEATION TEST REPORT
| Innovative Designs, Inc. | Filename: | J513T01 | ||||
| 124 Cherry Street, Suite 1 | Method Standard: | ASTM F739-99a | ||||
| Pittsburgh, PA 15223-2243 | Material Log No.: | 04513-88-01 | ||||
| Test Date: | 10/25/2005 | |||||
| JR KAV | ||||||
| Analyst QC | ||||||
| Material Name: | Camouflage Material | Test Temperature: | 27 deg C | |||
| Conditioning: | As Supplied | Test Duration: | 8 Hours | |||
| Chemical: | Gas Stimulant | Concentration: | 100% | |||
| Source: | In House | CAS No: | N/A | |||
| TEST RESULTS |
CELL 1 | CELL 2 | CELL 3 | AVERAGE | STD. DEV | |||||
| Actual Breakthrough (time in minutes) |
>480 | >480 | >480 | >480 | 0 | |||||
| Normalized Breakthrough (time in minutes) |
>480 | >480 | >480 | >480 | 0 | |||||
| Maximum Rate Detected (ug/cm3*min) |
<0.001 | <0.001 | <0.001 | <0.001 | 0 | |||||
| Unit Area Weight (g/cm2) |
0.032 | 0.032 | 0.032 | 0.032 | 0 | |||||
| Sample thickness (.001”) |
45 | 45 | 45 | 45 | 0 |
| Modifications of Method: | 1” Cells Open Looped System | |
| Analytical Technique: | Continuous Photoionization Lamp | |
| Sampling Frequency: | Every 3 minutes for the duration of the test | |
| Chemical physical state: | Gas | |
| Type of chemical contact: | Continuous | |
| Collection medium: | (Dehumidified Air) | |
| Minimum detectable rate: | 0.001 ug/cm2*min | |
| Notes: |
| |
The testing herein is based upon accepted industry practice as well as the test method used. The testing was performed under laboratory conditions, not actual usage conditions. Test results reported herein do not apply to samples other than those tested. TRI/Austin, Inc. makes no warranties or guaranties concerning protection by this material and assumes no liability for use of this material with the biological or chemical agent. The user should determine the applicability of the test conditions when accessing suitability of the material for actual anticipated exposure. TRI/Austin, Inc. neither accepts responsibility for nor makes claim as to the final use and purpose of the material. TRI/Austin, Inc. observes and maintains client confidentiality and limits reproduction of this report, except in full, without prior approval of TRI/Austin, Inc.
HYDROSTATIC RESISTANCE TEST REPORT
| Innovative Designs, Inc. | Filename: | J513T02 | ||||
| 124 Cherry Street, Suite 1 | Method Standard: | ASTM D751-95 | ||||
| Pittsburgh, PA 15223-2243 | Material Log No.: | 04513-88-01 | ||||
| Test Date: | 10/25/2005 | |||||
| JR KAV | ||||||
| Analyst QC | ||||||
| Camouflage Material | ||||||
| Material Name: | (Outer Shell Only) | Test Temperature: | 69.8 deg F | |||
| Conditioning: | As Supplied | Test Humidity: | 29% | |||
| Machine: | TRI-49 | No. of Replicates: | 10 | |||
| Source: | In House | Average Burst Strength: (psi) | 165.5 | |||
| Replicate # |
Burst Strength (psi) | |
| 1 |
160 | |
| 2 |
165 | |
| 3 |
165 | |
| 4 |
165 | |
| 5 |
170 | |
| 6 |
165 | |
| 7 |
165 | |
| 8 |
165 | |
| 9 |
165 | |
| 10 |
170 |